Polaris Generators

Polaris Generator Specifications

Parameters     

160kV

225kV

320kV

450kV

Output

DC Output Voltage

0-160kV

0-225kV

0 +/- 160kV

0 +/- 225kV

Max Output Current

30mA

30mA

30mA

30mA

Max Output Power

3kW

3kW

4.2kW

4.2kW

Polarity

Negative

Negative

Bi-Polar

Bi-Polar

Efficiency

>85%

>85%

>85%

>85%

Ripple/Noise (peak-to-peak)

0.1% (160Vpp)

.015% (338Vpp)

0.1% (320Vpp)

0.15% (675Vpp)

Input

Voltage Range

230VAC
(+10/-15%)

230VAC
(+10/-15%)

230VAC
(+10/-15%)

230VAC
(+10/-15%)

Current Supply

25A

25A

25A

25A

Control AC Input Current

6A

6A

6A

6A

# Phases

1

1

1

1

Line Frequency

47-63Hz

47-63Hz

47-63Hz

47-63Hz

Customer Interface

HV

R-24 Cable

R-28

2x R-24

2x , R-28

Remote Control Interface

Analog/RS232

Analog/RS232

Analog/RS232

Analog/RS232

Control AC Input

IEC Inlet

IEC Inlet

IEC Inlet

IEC Inlet

Mechanical/Environmental

Size

19"W x
10.5"H x
24"D

19"W x
17.5"H x
31.5"D

2x (19"W x
10.5"H x
24"D)

2x (19"W x
17.5"H x
31.5"D)

Weight

150 lbs

250 lbs

270 lbs

475 lbs

Operating Temperature

0° to 40°C

0° to 40°C

0° to 40°C

0° to 40°C

Humidity

5 to 95%
non-condensing

5 to 95%
non-condensing

5 to 95%
non-condensing

5 to 95%
non-condensing

Storage Temperature

-20° to 85°C

-20° to 85°C

-20° to 85°C

-20° to 85°C

*All high voltage connectors are tapered with flanged fittings

 

©2008 KIMTRON Inc.
© 2008 KIMTRON MEDICAL Ilc.